Case Series
| ||||||
Two different endotracheal tube manufacturing defects | ||||||
Ayse Belin Ozer1, Omer Lutfi Erhan2 | ||||||
1MD, Assistant Professor, Department of Anaesthesiology and Reanimation, Faculty of Medicine, Firat University, Elazig, Turkey.
2MD, Professor, Department of Anaesthesiology and Reanimation, Faculty of Medicine, Firat University, Elazig, Turkey. | ||||||
| ||||||
[HTML Full Text]
[PDF Full Text]
[Print This Article]
[Similar article in Pumed] [Similar article in Google Scholar] ![]() |
How to cite this article: |
Ozer AB, Erhan OL. Two different endotracheal tube-manufacturing defects. International Journal of Case Reports and Images 2013;4(3):150–153. |
Abstract
|
Introduction:
Although manufacturing defects related to airway equipment are not so common in practice, it risks the safety of the airway. Prompt recognition of equipment malfunction can prevent life threatening complications.
Case Series: Case 1: A 25-year-old patient had anesthesia induction for septorhinoplasty surgery. After the patient was ventilated via mask and intubated without problem, peak and plateau pressures increased above 35 cm H2O on controlled ventilation. It was noticed that the anesthetic reservoir bag did not maintain adequate compliance in a setting of high peak and plateau pressure and was thought to be due to a faulty endotracheat tube (ETT) connecter. We noticed that lumen of ET tube connector was narrow and connecter has been changed. After this change the relaxation of the balloon and a decrease in the ventilator pressures were noted. Case 2: A 45-year-old female patient had anesthesia induction for hysterectomy. After induction it was noted that there was a leak in the circuit, which was traced to the connection between the ETT and the inflation tube. An opening at the point of the connection of the EET with the inflation tube was detected. ETT was changed and anesthesia was maintained with no further problems. Conclusion: Difficult ventilation after successful endotracheal intubation can be due to equipment failure such as faulty ETT connecter and faulty inflation tube. | |
Keywords:
Endotracheal tube, Manufacturing defect, Leakage, Increased pressure
|
[HTML Full Text]
[PDF Full Text]
|
Author Contributions:
Ayse Belin Ozer – Substantial contributions to conception and design, Acquisition of data, Drafting the article, Revising it critically for important intellectual content; Final approval of the version to be published Omer Lutfi Erhan – Substantial contributions to conception and design, Acquisition of data, Drafting the article, Revising it critically for important intellectual content, Final approval of the version to be published |
Guarantor of submission
The corresponding author is the guarantor of submission. |
Source of support
None |
Conflict of interest
Authors declare no conflict of interest. |
Copyright
© Ayse Belin Ozer et al. 2013; This article is distributed the terms of Creative Commons Attribution License which permits unrestricted use, distribution and reproduction in any means provided the original authors and original publisher are properly credited. (Please see Copyright Policy for more information.) |
|